22
4.4 Pairing the remote control with SIMEOX
After pressing this button, the remote control and SIMEOX are paired, until SIMEOX returns to standby mode. This pairing operation therefore has to be repeated:
Every time SIMEOX is turned on Between two treatment sessions
When SIMEOX recognises the remote control, the CYCLE lights illuminate successively clockwise: the user is informed that the pairing has been performed successfully.
It is therefore normal that at the beginning of a session SIMEOX does not respond when the remote control button is first pressed, as this pressing is used to pair the two devices.
If the wireless connection between the remote control and the SIMEOX is interrupted, the CYCLE lights illuminate consecutively in a counter-clockwise direction: the user is informed that the remote control is un-paired.
The remote control is paired with SIMEOX simply by pressing once the remote control button after SIMEOX is activated and the expiratory kit has been recognised (light 1 steady blue, see above).