42
DIRECTIVES AND DECLARATION OF THE MANUFACTURER - EMISSIONS AND IMMUNITY
Emission tests Compliance
RF Emissions CISPR 11 Group 1
RF Emissions CISPR 11 Class B
MANUFACTURER S DECLARATION IMMUNITY TO RADIATED DISTURBANCES
Phenomenon Applied standard Immunity level
Magnetic field at mains frequency - (50/60 Hz) IEC 61000-4-8
30 A/m 50Hz
Radiated RF disturbances IEC 61000-4-3 80 MHz-2.7GHz Home care environment: 10 V/m 80% / 1 kHz
Radiated electromagnetic fields at radio wave frequencies IEC 61000-4-3
380 - 390 MHz - 27 V/m; PM 50%; 18 Hz 430 - 470 MHz - 28 V/m; (FM ±5 kHz, 1 kHz sine) PM; 18 Hz 704 - 787 MHz - 9 V/m; PM 50%; 217 Hz 800 - 960 MHz - 28 V/m; PM 50%; 18 Hz 1700 - 1990 MHz - 28 V/m; PM 50%; 217 Hz 2400 - 2570 MHz - 28 V/m; PM 50%; 217 Hz 5100 - 5800 MHz - 9 V/m; PM 50%; 217 Hz
Electrostatic Discharges (ESD) IEC 61000-4-2 Contact : ±8 kV - Air : ±15 kV
MANUFACTURER S DECLARATION IMMUNITY TO CONDUCTED DISTURBANCES
Phenomenon Applied standard Immunity level
Conducted disturbances caused by radio frequency fields
IEC 61000-4-6 150 kHz-80 MHz 3 V ISM/Radio frequencies 6 V 80% / 1kHz
Fast transient burst IEC 61000-4-4 5/50 ns, 100 kHz; ±2 kV
Surges IEC 61000-4-5 1.2/50 (8/20) μs line to line: ±1.0 kV
Voltage dips, short interruptions and voltage variations on power supply input lines IEC 61000-4-11
0% UT for 0.5 cycle (1 phase) 0% UT for 1 cycle 70% UT for 25/30 cycles (50/60 Hz)
Voltage dips, short interruptions and voltage variations on power supply input lines IEC 61000-4-11 0% UT for 250/300 cycles (50/60 Hz)
UT is the AC mains voltage prior to application of the test level