DIRECTIVES AND DECLARATION OF THE MANUFACTURER - EMISSIONS AND IMMUNITY

Emission tests Compliance

RF Emissions CISPR 11 Group 1

RF Emissions CISPR 11 Class B

MANUFACTURER S DECLARATION IMMUNITY TO RADIATED DISTURBANCES

Phenomenon Applied standard Immunity level

Magnetic field at mains frequency - (50/60 Hz) IEC 61000-4-8

30 A/m 50Hz

Radiated RF disturbances IEC 61000-4-3 80 MHz-2.7GHz Home care environment: 10 V/m 80% / 1 kHz

Radiated electromagnetic fields at radio wave frequencies IEC 61000-4-3

380 - 390 MHz - 27 V/m; PM 50%; 18 Hz 430 - 470 MHz - 28 V/m; (FM ±5 kHz, 1 kHz sine) PM; 18 Hz 704 - 787 MHz - 9 V/m; PM 50%; 217 Hz 800 - 960 MHz - 28 V/m; PM 50%; 18 Hz 1700 - 1990 MHz - 28 V/m; PM 50%; 217 Hz 2400 - 2570 MHz - 28 V/m; PM 50%; 217 Hz 5100 - 5800 MHz - 9 V/m; PM 50%; 217 Hz

Electrostatic Discharges (ESD) IEC 61000-4-2 Contact : ±8 kV - Air : ±15 kV

MANUFACTURER S DECLARATION IMMUNITY TO CONDUCTED DISTURBANCES

Phenomenon Applied standard Immunity level

Conducted disturbances caused by radio frequency fields

IEC 61000-4-6 150 kHz-80 MHz 3 V ISM/Radio frequencies 6 V 80% / 1kHz

Fast transient burst IEC 61000-4-4 5/50 ns, 100 kHz; ±2 kV

Surges IEC 61000-4-5 1.2/50 (8/20) μs line to line: ±1.0 kV

Voltage dips, short interruptions and voltage variations on power supply input lines IEC 61000-4-11

0% UT for 0.5 cycle (1 phase) 0% UT for 1 cycle 70% UT for 25/30 cycles (50/60 Hz)

Voltage dips, short interruptions and voltage variations on power supply input lines IEC 61000-4-11 0% UT for 250/300 cycles (50/60 Hz)

UT is the AC mains voltage prior to application of the test level