2.1 Alerts
Connect the device to a wall socket that supplies
the nominal voltage indicated on the machine.
In order to avoid any risk of arcing, connect the
power cord firstly to SIMEOX then to the wall
socket.
Likewise, once the device is switched off,
disconnect the cord from the wall socket before
disconnecting it from SIMEOX.
In order to avoid any risk of electrical discharge,
never plug in or unplug the device with wet hands,
do not use SIMEOX in damp rooms or in bathrooms.
Monitor the device during use and stop using it in
the event of a malfunction.
Do not let a young patient use SIMEOX without
adult supervision.
The power cable and the tube of the expiratory
kits present a strangulation risk: keep them out of
the reach of children.
The mouthpiece and the remote control battery
present a suffocation risk: keep them out of the
reach of children.
Only use the accessories and expiratory kits
supplied by PHYSIO-ASSIST.
SIMEOX is designed and manufactured to resist
spills of liquids falling vertically on the cover, but
as a safety measure, in the event of contact with
a liquid, immediately switch the device off and
disconnect it from the mains.
Never use SIMEOX if the plug or the power cord
is damaged, if the device presents an operating
fault or if it has been dropped, damaged or
immersed in water.
Never dismantle SIMEOX and do not try to
intervene if it breaks down: SIMEOX does not
contain any parts to be replaced by the user.
You are only responsible for replacing the remote
control batteries, as described in this manual.
Only approved personnel are authorised to
perform maintenance operations on SIMEOX.
Do not install SIMEOX against a wall; the main
switch must always remain accessible to switch
off SIMEOX in an emergency.
All modifications to SIMEOX are prohibited.
The male mains plug can be used as a means of
separating SIMEOX from the electricity grid.
The SIMEOX mouthpiece is intended to make
contact with the patient’s mouth.
Although the materials used are biocompatible,
in the event of irritation, tingling or allergies,
stop using SIMEOX immediately and consult a
specialist.
Before cleaning, always switch off the device and
unplug it from the wall socket.
When cleaning or disinfecting expiratory kits,
make sure to remove all product residues.