3.4 Control pad components
1
“Standby/On” button
This button activates operation of SIMEOX and its
“standby” mode.
When the light is yellow-orange, the device is in
standby.
When the light is blinking blue, the device is active
but is not yet usable as it has not detected a valid
accessory (usable SIMEOX expiratory kit): either
the expiratory kit used is not supplied by PHYSIO-
ASSIST, or the SIMEOX consumable allowed use
counter is depleted.
In this case: you must replace the accessory to be able
to use SIMEOX.
When the light is steady blue (no blinking), the
device is active and usable.
As SIMEOX is not watertight and does not withstand
being dropped, it is preferable to use the bag
provided to transport it. In this way your SIMEOX
will be best protected from water and impacts.
2
Program adjustment button
This button allows the program to be adjusted to
the patient according to their preference:
PROG 1: program 1 is selected for 6 expiratory
phases per cycle
PROG 2: program 2 is selected for 8 expiratory
phases per cycle
PROG 3: program 3 is selected for 10 expiratory
phases per cycle
It is possible to change the program at the end of each
cycle, or during the 45-second pause.
For each program: The last two phases have a
frequency that is lower than that of the preceding
phases, in order to increase transport of the mucus.
This button also triggers a drying tube cycle, see
below.
3
Power adjustment button
This button adjusts the power of the signal sent
to the patient. It can be adjusted at any time in
the session based on how the patient feels.
It is possible to change the power setting at
any time: during the session, at the end of each
program, or during the 45-second pause.
4
Remote control battery indicator
This light changes to orange when the remote
control battery needs to be replaced.
5
Performance indicator
This visual feedback enables the operating quality
of the device to be viewed.
The number of green lights illuminated (between 1
and 5) is an indicator of the quality of transmission
of the pneumatic signal in the lower airways
Three to five green lights should progressively
illuminate when the patient expires.
If the red lights illuminate, the expiration must be
stopped by releasing the remote control button
or removing the mouthpiece.
1
3
5
2
4
16
Tap twice or spread your fingers to zoom in
Close menus and pop-ups
Zoom in and zoom out
Click once to zoom in, click again to zoom out
Roll the mouse wheel to zoom in/out