10.5 Compatible accessories
10.5.1 Expiratory kits:
Single patient kit with disposable mouthpieces, 10
uses, in a box of 10 kits, reference: TUB10_EU
Single patient kit with reusable mouthpieces, 25
uses, in a box of 10 kits, reference: TUB25_EU
10.5.2 Accessories:
Carrying bag (supplied), reference SAC01
Wrist strap (supplied), reference DRAG01
10.6 Communications
Warning! Because of its Bluetooth, Bluetooth Low
Energy and RFID components, SIMEOX can be subject
to interference from other products, even if they
comply with CISPR emission requirements.
SIMEOX uses wireless communications in the following
bands:
To communicate via Bluetooth with a tablet,
SIMEOX uses GFSK modulation for an effective
radiated power of 3mW on the band [2.400 GHz-
2.4835 GHz].
To communicate via RFID with the consumable,
SIMEOX uses ASK modulation for an effective
radiated power of 0.2512μA/m in the frequency
13.56MHz.
To communicate with its remote control, SIMEOX
uses a Bluetooth BLE interface: [2.400 GHz-
2.4835 GHz] with a bandwidth of 2MHz per
communication channel.
10.7 Stipulated environmental
conditions
SIMEOX is intended to be used: in healthcare facilities,
in medical or paramedical practices, at home.
The conditions of use are the following (SIMEOX could
not operate in accordance with its specifications if
these are not present):
Temperature limits at the place of use: +5 to +40°C
Humidity limits at the place of use: 15% to 93%
Atmospheric pressure at the place of use: 700hPa
to 1060hPa
When SIMEOX is subjected to extreme temperatures
(low or high), make sure to wait at least 4 hours
before using, to allow its components to return to a
temperature that allows for its optimal use.
10.8 Electrical characteristics
10.8.1 Supply voltages:
SIMEOX_H_EU : 230 V AC 50 Hz
The SIMEOX is designed and manufactured
to operate within the supply voltage limits of
-15%/+10% set by European standards, that is 195
VAC to 253 VAC.
10.8.2 Applied parts (according to
standard EN 60601-1):
In compliance with standard EN 60601-1.
When used at home, the patient being the user, all
accessible parts of SIMEOX are considered applied
parts.
These are Type BF applied parts.