SYMPTOM PROBABLE CAUSE(S)ACTION TO BE UNDERTAKEN
SIMEOX no longer starts at all, the
orange light does not illuminate
1- Defective power supply
2- Power cord damaged
3- Possible failure of SIMEOX electronics
1- Check the power supply is functioning
correctly
2- Disconnect SIMEOX from the mains and
check the power supply cord.
3- If the symptom persists: possible failure
of the SIMEOX electronics
Contact PHYSIO-ASSIST or an approved
centre
SIMEOX starts but stops unexpectedly
Inappropriate power supply: SIMEOX is
intended to operate with a power supply
that conforms to standards, in the case of
a power supply that does not conform to
standards, or in the event of repeated short
power outages, SIMEOX switches to “safety”
mode. Wait a few minutes and try again.
Have your power supply checked by a
specialist.
If the supply complies with European
standards, and the symptom persists:
possible failure of the SIMEOX electronics.
Contact PHYSIO-ASSIST or an approved
centre
Compressor noise as soon as it is
turned on
Failure of the SIMEOX electronics
Contact PHYSIO-ASSIST or an approved
centre
SIMEOX does not recognise the
PHYSIO-ASSIST expiratory kits
The consumable is no longer usable: try with
a new consumable.
If the symptom persists: possible failure of
the SIMEOX electronics
Contact PHYSIO-ASSIST or an approved
centre
SIMEOX seems to operate but no
negative pressure pulse output
Internal failure of SIMEOX (electronics,
compressor, etc.)
Contact PHYSIO-ASSIST or an approved
centre
SIMEOX operates but the operating
noise is irregular and loud
Internal failure of SIMEOX (electronics,
compressor, etc.)
Contact PHYSIO-ASSIST or an approved
centre
SIMEOX no longer responds to the
remote control.
It is impossible to pair SIMEOX and the
remote control.
Replace the remote control battery.
If the symptom persists: possible failure of
the SIMEOX electronics or of the remote
control.
Contact PHYSIO-ASSIST or an approved
centre
When checking the lights on starting
the device, one light does not
illuminate
Control pad needs to be replaced
Contact PHYSIO-ASSIST or an approved
centre
In the event of malfunction, the user can however carry out simple manipulations to confirm that the fault is indeed
attributable to SIMEOX.
46