Some chemical products used as disinfectants
can harm the organism and the expiratory kits of
the device.
After cleaning and drying, store the device and
its accessories in its carrying bag, in a dry place,
protected from sunlight and dust, at room
temperature.Please refer to the section in this
manual on storage temperatures for SIMEOX.
The Lithium button battery in the remote control
presents an ingestion risk, to avoid this risk take
the following precautions:
- Do not store new batteries within the reach of
children
- Do not store used batteries in your home,
take them to a collection point
- Do not replace the batteries in the presence
of a child
- Do not let children play with the batteries
- Never put the battery in your mouth to test it
or to keep your hands free
- Never place the battery beside your
medicines, to avoid mistaking it for a tablet
Ingesting a button battery may cause the
following injuries:
- necrosis and perforation of the oesophagus
through mechanical compression
- electrical burns caused by the heat from the
electrical current between the mucosa and
the battery.
In the event that you ingest a button battery
contact a doctor or the emergency services
immediately.
The base of the SIMEOX housing incorporates
openings intended to aerate the device and to
evacuate any condensed moisture.
At home in particular, make sure that no objects,
especially metal objects such as staples,
paperclips, etc. are introduced into the housing
through these openings.
Although the live parts cannot be accessed
through these openings, a small object could
remain stuck in the device and cause a short-
circuit.
The wrist strap supplied is intended to ensure
that the remote control is held securely at the
wrist during treatment sessions, preventing it
from being dropped.
It is not pre-installed.
The user is free to install it if he chooses,
depending on the estimated contamination risk.
Keep the power cord away from all hot surfaces:
the heat could degrade the protective sheath and
cause a failure, or a risk of electric shock.
Do not position SIMEOX in such a way that it is
difficult to use the disconnection device (the male
mains plug).
10
Tap twice or spread your fingers to zoom in
Close menus and pop-ups
Zoom in and zoom out
Click once to zoom in, click again to zoom out
Roll the mouse wheel to zoom in/out