3.1 Components on the back panel
Description of SIMEOX
3
1
Power switch
This I/O switch turns the SIMEOX device on and off,
“I” to turn on and “O” to turn off.
2
Mains f ilter fuse
This location is intended for the mains filter fuses.
3
Power socket
This connector is for the SIMEOX C17 power cable.
1
2
3
Destruction of fuses can only be caused by a critical failure of the SIMEOX electronics, for example in the
event of exposure to very high electrostatic discharge, overvoltage, overcurrent, etc.
Tools are required to replace fuses and this must imperatively be carried out by an approved technician
who will first examine the device to determine the cause of destruction.
13