DIRECTIVES AND DECLARATION OF THE MANUFACTURER - EMISSIONS AND IMMUNITY
Emission tests Compliance
RF Emissions CISPR 11 Group 1
RF Emissions CISPR 11 Class B
MANUFACTURER’S DECLARATION – IMMUNITY TO RADIATED DISTURBANCES
Phenomenon Applied standard Immunity level
Magnetic field at mains frequency -
(50/60 Hz)
IEC 61000-4-8
30 A/m
50Hz
Radiated RF disturbances IEC 61000-4-3
80 MHz-2.7GHz
Home care environment: 10 V/m
80% / 1 kHz
Radiated electromagnetic fields at radio
wave frequencies
IEC 61000-4-3
380 - 390 MHz - 27 V/m; PM 50%; 18 Hz
430 - 470 MHz - 28 V/m; (FM ±5 kHz, 1 kHz sine) PM; 18 Hz
704 - 787 MHz - 9 V/m; PM 50%; 217 Hz
800 - 960 MHz - 28 V/m; PM 50%; 18 Hz
1700 - 1990 MHz - 28 V/m; PM 50%; 217 Hz
2400 - 2570 MHz - 28 V/m; PM 50%; 217 Hz
5100 - 5800 MHz - 9 V/m; PM 50%; 217 Hz
Electrostatic Discharges (ESD)IEC 61000-4-2 Contact : ±8 kV - Air : ±15 kV
MANUFACTURER’S DECLARATION – IMMUNITY TO CONDUCTED DISTURBANCES
Phenomenon Applied standard Immunity level
Conducted disturbances caused by radio
frequency fields
IEC 61000-4-6
150 kHz-80 MHz 3 V
ISM/Radio frequencies 6 V
80% / 1kHz
Fast transient burst IEC 61000-4-4 5/50 ns, 100 kHz; ±2 kV
Surges IEC 61000-4-5 1.2/50 (8/20) μs line to line: ±1.0 kV
Voltage dips, short interruptions and voltage
variations on power supply input lines
IEC 61000-4-11
0% UT for 0.5 cycle (1 phase)
0% UT for 1 cycle
70% UT for 25/30 cycles
(50/60 Hz)
Voltage dips, short interruptions and voltage
variations on power supply input lines
IEC 61000-4-11 0% UT for 250/300 cycles (50/60 Hz)
UT is the AC mains voltage prior to application of the test level